Hardware test day

The written procedure for the hardware tests (design §10). Each tier needs the owner's authorization before it is ever run. This page covers:

  • the read-only tier;
  • the stateful tier, which writes settings and restores them;
  • the watching of a calibration the owner makes at the front panel;
  • the key prototype, which presses front-panel keys over Modbus without calibrating.

There are no destructive tests: the bench analyzer cannot auto-calibrate. Results go to protocol-findings.md.

Bench facts

Analyzer ZPA, serial N8A0259T, program version 1.02
Channels Ch1 CO2 0-10 vol%, Ch2 CO 0-1 vol%, Ch3 O2 0-21 / 0-25 vol%
Link RS-485 through an FTDI FT232R adapter; COM8 when last checked
Station 1
Framing 38400 8-N-1, fixed
Firmware features clock and A/D block present; no calibration log and no type-code digits 27-29

COM port numbers can change when adapters are replugged. When in doubt, find the port by unplugging and replugging the analyzer's adapter while listing the ports. That sends nothing to the other instruments on the rig.

Scope of the read-only session

Allowed Not allowed
Every read the library makes: identity, polls, status, metadata, settings, logs, clock, A/D Any register write
Discovery on the analyzer's own port, its station and the next one Operation commands: auto calibration, auto zero, blowback, return to measurement
The command-line tools, which only read Key simulation (design §6.5, Phase 7B)
Probing ports that belong to other instruments

How the bench is driven on the development machine

Serial work on the development machine has been run from Git Bash, with uv on the path:

cd fujilib
export PATH="$HOME/.local/bin:$PATH"
FUJILIB_ENABLE_HARDWARE_TESTS=1 FUJILIB_HARDWARE_PORT=COM8 \
    uv run pytest -m hardware tests/hardware -v

FUJILIB_HARDWARE_ADDRESS sets the station when it is not 1. The tests run under asyncio and trio.

Pre-flight checklist

  1. The analyzer is powered on and shows the measurement screen.
  2. No one is working at the analyzer's front panel. The panel stays live, and a range changed there shows up in the tests.
  3. Nothing else has the port open: no terminal program, no other session.
  4. The port is the analyzer's (see Bench facts).

The read-only session

tests/hardware/ holds four files:

File Covers
test_hardware_client.py the transport, the Modbus client and the read procedures: identity, poll timing after the inter-frame gap, status, ranges, metadata, settings, logs, clock and A/D, 50 sustained polls, and a cancelled read that must not disturb the next
test_hardware_reads.py the Analyzer facade: open and identify, an asserted label, a poll of exactly two transactions, polls without detail, status, metadata, ranges, settings and parameters, the logs (the calibration log refused before any I/O on firmware 1.02), clock, A/D and reprobing, 50 sustained polls, a deadline, an empty station that times out and releases the port, and closing and opening again
test_hardware_sync.py the blocking facade, discovery on the analyzer's port, and fuji-read, fuji-discover and fuji-configure dump
test_hardware_recording.py a 5 Hz recording (schedule, fixed row columns), pipe() to CSV and Parquet, fuji-stream, fuji-capture, and a short fuji-diag timing run

Every assertion holds for any ZP analyzer; values particular to the bench unit are recorded in the findings instead.

The 12-hour recording

The hardware exit of Phase 5 (design §12, §13.1 #58). It is read-only, and it holds the port for 12 hours; overnight suits it.

  1. Pre-flight as above, plus: the host does not sleep, USB selective suspend is off for the adapter, and no restart for updates is due.
  2. Open a console window yourself (Command Prompt, PowerShell or Windows Terminal), go to the repository, and start it on one line:
uv run --with psutil python scripts/soak_monitor.py --log probe_out/soak.rss.jsonl -- fuji-capture COM8 --gas CH1=co2 --gas CH2=co --gas CH3=o2 --rate 1 --duration 43200 --out probe_out/soak.parquet --reconnect

A progress line every 10 s shows the polls, failures and late ticks so far; the memory log gets a line every 10 minutes, and probe_out/soak.parquet.meta.json gets the counters so far every minute. Selecting text in the window pauses the progress line until the selection ends, but not the recording. 3. Leave it alone. A front-panel change is recorded, not an error. 4. To stop it early, press Ctrl-C; if Ctrl-C does nothing, press Ctrl-Break. Either stops it cleanly: wait for its summary before closing the window. Never close the window while it records: that kills it, and a killed Parquet file has no footer and cannot be read. 5. When it ends, check it:

uv run python scripts/check_soak.py probe_out/soak.parquet --rss probe_out/soak.rss.jsonl

It checks tick and row counts, timing, error accounting, status and provenance in every row, a clean shutdown, readable output and bounded memory (the trend of private memory after the first hour), and writes probe_out/soak.parquet.check.json. 6. If it was killed anyway, recover the complete row groups (1,000 rows each) into probe_out/soak.recovered.parquet, with a copy of the .meta.json beside it, and check that file instead. It fails the shutdown check, and the counts it cannot judge without final counters are marked SKIP:

uv run python scripts/recover_parquet.py probe_out/soak.parquet
uv run python scripts/check_soak.py probe_out/soak.recovered.parquet --rss probe_out/soak.rss.jsonl

The unplug test

A controlled disconnect and reconnect (design §12). Someone must be at the bench to pull the adapter's USB plug. Read-only.

  1. Start a 10-minute capture with reconnection:
fuji-capture COM8 --gas CH1=co2 --gas CH2=co --gas CH3=o2 \
    --rate 1 --duration 600 --out probe_out/unplug.csv --reconnect
  1. After about two minutes, pull the adapter's USB plug; put it back after about 30 s. Repeat once. The progress line counts the failed polls meanwhile.
  2. Expected: the capture finishes with exit code 0; the rows of each outage are error rows (FujiConnectionError, then refusals until the port is back); probe_out/unplug.csv.meta.json reports 2 disconnects and 2 reconnects. Check that the port came back as COM8.
  3. Run it again without --reconnect: at the first pull it ends with exit code 1, and the CSV and its .meta.json (state failed) are complete up to the failure, with 1 disconnect and 0 reconnects.

The stateful session

It writes settings, reads each back and restores it, and needs its own authorization and the owner at the front panel. Nothing in it starts a calibration: the bench analyzer's type code lists no auto-calibration option, and the analyzer has no calibration valves of its own, so auto calibration would calibrate on whatever gas is at the inlet (Safety).

Allowed Not allowed
Writes of the reviewed settings, each restored Auto calibration, auto zero calibration (never on this analyzer)
Return to measurement Blowback (not a ZPA feature)
scripts/probe_write.py, step by step Key simulation (design §6.5, Phase 7B)
Any register outside the reviewed subset
  1. Pre-flight as for the read-only session, and save the settings first:
fuji-configure dump COM8 --out probe_out/settings_before.json
  1. The stateful tests. The first ones write only registers of channels and NDIR components the bench analyzer does not have (Ch4, Ch5, NDIR 4), with both FC06 and FC10; then the O2 response time, output hold, hold mode and Ch3's range, each restored; then a settings document and its baseline. A fixture reads every writable setting before each test and checks after it that the analyzer has them all back.
FUJILIB_ENABLE_STATEFUL_TESTS=1 FUJILIB_HARDWARE_PORT=COM8 \
    uv run pytest -m hardware_stateful tests/hardware -v
  1. Key lock (design §13.2 #12). Switch key lock on at the front panel, then run the step. It writes one register of Ch5 and sends return to measurement, records whether each was applied, ignored or refused, and restores the register. Switch key lock off at the panel afterwards.
uv run python scripts/probe_write.py --port COM8 key-lock --confirm
  1. A menu at the panel. Open any menu at the front panel, then try a write (fuji-configure apply of a one-setting document): it must be refused with nothing written. Then return_to_measurement(confirm=True) must bring the panel back to the measurement screen.
  2. Power cycle (design §13.2 #14). Write a marker, switch the analyzer off and on, wait for the measurement screen, and check it; the second step restores the register.
uv run python scripts/probe_write.py --port COM8 persist-write --confirm
# switch the analyzer off, then on
uv run python scripts/probe_write.py --port COM8 persist-check --confirm
  1. Values out of range, if authorized on the day. It writes 61, then 0, to the response time of NDIR component 4 through the client, past the library's own limit, records whether the analyzer refuses, clamps or stores each, and restores it.
uv run python scripts/probe_write.py --port COM8 out-of-range --confirm
  1. The start time of the schedules (design §13.2 #26): read the auto-calibration start time the front panel shows. No Modbus is needed.
  2. Close: compare with the saved settings; nothing may be written.
fuji-configure diff COM8 --file probe_out/settings_before.json

Every setting must be unchanged. If one is not, apply the file with fuji-configure apply ... --confirm (and the destructive flag for a calibration gas).

Stop at the first unexpected result and record it: a write that reads back otherwise, an unknown outcome, or a setting left changed.

Watching a calibration at the panel

fujilib only reads here. The owner calibrates at the front panel, with the gases at the inlet, and decides each step; it needs the owner's authorization like any session, because it changes the analyzer's calibration. It checks that every manual zero and span is recorded as it happened (design §6.5, Phase 7A).

  1. Pre-flight as for the read-only session. Note the calibration gas settings of the channel to be calibrated (fuji-configure dump), and whether it is set to zero "at once" or calibrate "both" ranges: plan_manual_calibration() lists what a zero or span would touch.
  2. Start the watcher, and leave it running:
uv run python examples/watch_manual_calibration.py COM8
  1. A zero, with zero gas flowing: ZERO, the cursor to the channel, ENT to select it, and ENT again once the reading is steady. Expected: one line with "outcome": "completed", the channel, the reading before, after at the zero gas, and the deviation.
  2. A span, with span gas flowing: SPAN, the channel, ENT, ENT. Expected as for the zero, with after at the span gas setting.
  3. A cancel: ZERO, the channel, ENT, then ESC. Nothing is calibrated. Expected: "outcome": "cancelled".
  4. Stop the watcher with Ctrl-C.

The 2026-09-29 sessions recorded the same sequence with scripts/probe_calibration.py (findings §14) and then with this watcher, which reported the zero and span completed and the cancel cancelled (findings §16). The key prototype below answered which flags the "at once" pair sets (findings §18.5). A channel set to "both", output hold and a calibration that fails are still open (design §13.2 #36-#38).

The key prototype

This session finds out what a key written to 42001 does before fujilib drives a manual calibration itself (design §6.5, §12 Phase 7B). scripts/probe_panel.py presses front-panel keys over Modbus, one experiment per run. It needs the owner's authorization for the session, with the owner at the front panel throughout (design §13.1 #79). Nothing in it calibrates.

Allowed Not allowed
scripts/probe_panel.py, one experiment at a time MODE and SIDE, which open the menus and enter their passwords; two keys at once
ZERO, SPAN, UP, DOWN and ESC; ENT on channel selection, where it selects the channel ENT on a wait step, where it starts the calibration, or on the error display, where it can force one (ZPA p.89)
42002, return to measurement 42003-42005: auto calibration, auto zero, blowback
Output hold switched on for the hold step and off after it, with fuji-configure apply Any calibration. The error display and a channel set to "both" need one (design §13.2 #36, #37), so they are left out
Any other register write

The probe enforces this itself:

  • Its one write takes an address and a value from a fixed list: the six keys to 42001, and 1 to 42002.
  • It reads the panel just before every key, and refuses a key the step does not allow. ENT also needs the cursor on the channel the experiment names.
  • A key is sent once and never repeated. A lost reply is settled by reading the panel.
  • After each key it reads the step, the cursor, the zero, span and hold flags, 00B9h and 00BDh until they settle, and records when each changed.
  • Anything unexpected ends the experiment, and a cleanup follows:
  • ESC on channel selection, a wait step or the error display;
  • a wait while a calibration runs;
  • 42002 on any other screen.

It then checks the flags, not only the screen. It sends each of its keys once, and not at all if the experiment has just sent the same key on the same step.

At the panel:

  • Keep your hands off the keys while an experiment runs. The probe reads the step just before each key, so a key pressed at the panel between that read and the write lands on a step it did not check.
  • If it is practical, have zero gas at the inlet for the zero experiments and span gas for the span ones. Then even a calibration started by mistake would be made on the right gas. The probe never sends the ENT that calibrates, so this is a second line of defence, not a requirement.
  • Say what the display shows where the probe cannot see it: the backlight, a key-lock message.

Before the first key

  1. Pre-flight as for the read-only session. Check the process list: no recording, soak monitor or probe has COM8 open.
  2. Check the probe on the simulator. It opens no port and must end with every check passed:
uv run python scripts/probe_panel.py check
  1. Save the settings:
fuji-configure dump COM8 --out probe_out/settings_before_keys.json
  1. Read the panel and the settings the experiments depend on. This reads only:
uv run python scripts/probe_panel.py --port COM8 status

Expected: - the measurement screen, step 0, no calibration or hold flag set; - key lock off and output hold off; - Ch3 zeroed on its own ("each"), and Ch1 and Ch2 zeroed "at once".

The experiments

Run each in this order, one at a time:

uv run python scripts/probe_panel.py --port COM8 <experiment> --confirm
# Experiment Keys sent Expected What it answers
1 select-esc ZERO, ESC; SPAN, ESC step 0 → 4 → 0, then 0 → 7 → 0; no flag set whether a key acts, whether it shows in 00BDh, how soon the step follows; ESC from channel selection
2 cursor ZERO, DOWN ×3, UP ×3, ESC; then the same after SPAN a zero offers the Ch1 position (Ch1 and Ch2 "at once") and Ch3; a span offers Ch1, Ch2 and Ch3 whether the cursor stops at the ends or wraps round
3 zero-cancel ZERO, UP or DOWN to Ch3, ENT, ESC step 5 with Ch3's zero flag only, 00B9h at 0; ESC gives step 0 with the flag clear the channel-selecting ENT, and ESC from the wait step
4 span-cancel SPAN, to Ch3, ENT, ESC step 8 with Ch3's span flag only; ESC gives step 0 with the flag clear the same for a span
5 return-select ZERO, 42002 not known what 42002 does on channel selection
6 return-wait ZERO, to Ch3, ENT, 42002 the measurement screen with Ch3's zero flag left set (findings §18.4); clear it at the panel: ZERO, O2, one ENT, ESC what 42002 does on a wait step, and whether the flag clears with the step
7 at-once ZERO, UP to the Ch1 position, ENT, ESC step 5 with the zero flags of Ch1 and Ch2, perhaps also of the absent Ch4 and Ch5; ESC clears them all design §13.2 #37, the "at once" pair
8 key-lock ZERO, then ESC if it opened channel selection not known whether key lock stops a key written over Modbus
9 backlight ZERO, then ESC if it opened channel selection not known whether a key is swallowed while the backlight is off
10 hold ZERO, to Ch3, ENT; 120 s on the wait step; ESC see below design §13.2 #38

Key lock (8). Switch key lock on at the front panel (Parameter Setting), return to the measurement screen, and run the step. Say what the display does. Switch key lock off afterwards.

Backlight (9). Let the backlight go off. Its timer is set in Parameter Setting, 1-60 minutes; set it as you prefer, and put it back afterwards. Run the step without touching the panel, and say whether the backlight came on.

Output hold (10). Switch output hold on with a one-setting document, then run the step in the background: it takes up to 10 minutes.

fuji-configure apply COM8 --file probe_out/output_hold_on.json --confirm
uv run python scripts/probe_panel.py --port COM8 hold --confirm

output_hold_on.json is {"format": "fujilib-settings/1", "settings": {"output_hold.enabled": true}}.

  1. Start with span gas (air) at the inlet.
  2. When the panel shows the zero wait screen for Ch3, switch to zero gas. The probe reads the readings and the A/D values for 120 s (--dwell), then sends ESC.
  3. After ESC it reads on until the hold flag clears, for at most 7 minutes (--hold-watch). After a calibration is cancelled, the analyzer holds the outputs for the hold extension (ZPA p.66 e.), and this unit's gas flow times are all 300 s.
  4. Keep the zero gas until the probe ends, then switch back to air.
  5. Switch output hold off with the same document set to false.

Expected (ZPA p.67): the Modbus readings stay at their value from before the hold, while the O2 A/D count follows the gas. That would let the steadiness rule use the counts where hold is on (design §13.1 #78). To end the watching early, create probe_out/probe_panel.stop: the probe sends the ESC and stops.

Closing

Compare with the saved settings; nothing may be written. Then read the panel:

fuji-configure diff COM8 --file probe_out/settings_before_keys.json
uv run python scripts/probe_panel.py --port COM8 status

Every setting must be unchanged, and the panel on measurement with no flag set.

When something is unexpected

  • An experiment that ends unexpected, or a cleanup that ends NOT CLEAN, stops the session. Look at the panel. Run nothing more until status shows the measurement screen with no flag set.
  • If a flag stays set on the measurement screen, the probe sends nothing more, and the owner decides at the panel. As a last resort, switching the analyzer off and on clears the panel's state (findings §15.5); the readings are wrong for about a minute afterwards (design §13.1 #81).

Not in this session:

  • the error display (design §13.2 #36) and a channel set to "both" (#37). Both need a real calibration, and one that fails can change the calibration. They need the owner's explicit acceptance, in a session of their own;
  • anything touching auto calibration, auto zero or blowback, which are never run on this unit.

Each run writes probe_out/probe_panel_<experiment>_<time>.json. The results go to protocol findings §18.

The session of 2026-09-30 (12:56-13:12 UTC, findings §18) ran experiments 1-8. span-cancel came after at-once, so the gas changed only once. The backlight and output-hold steps were not run. Every key acted as at the panel, and nothing was calibrated. At the close all 162 settings matched the saved ones. Only return-wait did not end clean: its 42002 left Ch3's zero flag set, and the owner cleared it at the panel.

Remote zero and span

This session makes a manual zero and a span of O2 from the host (design §6.5, §12 Phase 7C): fuji-calibrate presses the calibration keys, the owner switches the gas valves. It changes the analyzer's calibration. It needs the owner's authorization for the session, with the owner at the gases and the panel throughout (design §13.1 #79), zero gas (nitrogen) and span gas.

Allowed Not allowed
fuji-calibrate, one run at a time, on Ch3 (O2) MODE, SIDE, or any key but the six of a manual zero or span (the write envelope refuses them anyway)
A zero of Ch3 on zero gas, a span of Ch3 on span gas, and a cancel A zero of the "at once" pair (CO2 and CO) unless the owner asks for it: it calibrates both
Reading the factory Coefficient screen after the zero, at the panel Changing a calibration-gas setting in the session, unless the owner decides to

fuji-calibrate enforces its part itself: it refuses before the first key while key lock or output hold is on, a flag is set or the gas named is not the calibration-gas setting; it asks before the key that calibrates; and it returns the panel to measurement however the run ends.

Before the first key

  1. Pre-flight as for the read-only session. Check the process list: no recording, soak monitor or probe has COM8 open.
  2. Run it on the simulator. It opens no port:
fuji-calibrate --fixture bench --channel CH3 --kind span --gas-value 20.95 --gas-unit vol% --confirm --i-understand-this-is-destructive --auto --window 2 --response-factor 0 --out probe_out/fixture_span.json

Expected: status: completed. 3. Save the settings, and note Ch3's range-1 calibration gases (0.00 and 20.95 vol% on 2026-09-29):

fuji-configure dump COM8 --out probe_out/settings_before_remote_cal.json
  1. The plan, which reads only:
fuji-calibrate COM8 --gas CH1=co2 --gas CH2=co --gas CH3=o2 --channel CH3 --kind zero --plan

Expected: CH3: range 1 against 0 vol%, no other channel, no "at once" note. 5. Optional, before the calibrations: the output-hold step of the key prototype (probe_panel.py hold, design §13.1 #86 (a)) and its backlight step (§13.2 #39), as that session describes them.

The runs

fuji-calibrate prints the reading as it settles, asks before it calibrates, and ends with status:. Each run writes probe_out/remote_<kind>_<n>.json with --out.

  1. A cancel. Zero gas at the inlet, then:
fuji-calibrate COM8 --gas CH1=co2 --gas CH2=co --gas CH3=o2 --channel CH3 --kind zero --gas-value 0 --gas-label "N2" --confirm --i-understand-this-is-destructive --out probe_out/remote_zero_1.json

When it asks, answer n. Expected: status: cancelled, the panel on measurement, and nothing calibrated (00B9h stays 0). 2. A zero. The same command with --out probe_out/remote_zero_2.json. Keep zero gas flowing; answer y once it says steady. Expected: status: completed, O2 at 0.00 after, a deviation of a few hundredths, the O2 detector near 634 counts (findings §16). 3. Optional: read the factory Coefficient screen for Ch3 range 1 at the panel (findings §17.3), and note the zero coefficient with the zero's count from the record. 4. A span. Switch to span gas, then:

fuji-calibrate COM8 --gas CH1=co2 --gas CH2=co --gas CH3=o2 --channel CH3 --kind span --gas-value 20.95 --gas-unit vol% --gas-label "span gas" --confirm --i-understand-this-is-destructive --out probe_out/remote_span_1.json

Answer y once steady. Expected: status: completed, O2 at 20.95 after. 5. Switch the inlet back to the sample.

Closing

fuji-configure diff COM8 --file probe_out/settings_before_remote_cal.json
fuji-read COM8 --include status

Every setting must be unchanged, and the panel on measurement with no flag set.

When something is unexpected

  • A run that ends not_clean, or raises that a flag is still set, stops the session. The message names the channel; recover at the panel: ZERO or SPAN, the channel, ENT once, then ESC.
  • A run that ends refused sent nothing; its message says why (key lock, output hold, a flag, the gas named).
  • If the reading never settles within 10 minutes, the run ends cancelled. The records keep every read on the wait step (wait_series), from which the steadiness rule's defaults are to be tuned (design §13.1 #78).

The results go to protocol findings §19, and the tuned defaults to design §13.1 #78.

The session of 2026-09-30 (15:32–15:36 UTC, findings §19) ran the cancel, the zero and the span, with the owner's answers passed on at the prompt. The zero took O2 from 0.05 to 0.00 vol% on N2 and the span from 20.89 to 20.95 vol% on air; the cancel ran nothing. Every cleanup found the panel clean and every setting matched the saved ones. The gases were already steady when each run began; a run started before the gas is switched would record how long it takes to settle. Neither optional step (output hold, the backlight) was run, and the Coefficient screen was not read.

Deliverables

  • The test run's summary, with any failures and whether they reproduce.
  • New findings, dated, in protocol-findings.md, and the design sections they change.
  • Captures, if any were made, kept in tests/fixtures/captures/ (git-ignored).

Risks

  • Background loss. About one request in 3,000 goes unanswered (findings §6.3). A read retries, so a test that counts attempts can fail on it; rerun that test on its own before calling it a fault.
  • Port release on Windows. Closing and reopening the same port at once has worked on the bench; if it ever fails with "access is denied", wait briefly and retry.
  • The operator. A front-panel change during the run (range, hold, calibration) changes what the tests read.