Async-first Python driver for Fuji Electric ZP-series NDIR gas analyzers: the ZPA, and the ZPB / ZPG / ZPAJ / ZPG3E models that share its MODBUS map. It is built on anyserial and anymodbus.

The ZPA is also sold as the CAI ZPA by California Analytical Instruments (CAI), now part of ENVEA. fujilib is developed and tested against a CAI-branded ZPA. See Also sold as.

Alpha

0.1.0 is the first release. The read-only API works against the development analyzer: open_device(), identification, polls with validity, metadata, settings, logs, discovery, recording to memory, CSV or Parquet, a blocking facade and the fuji-* commands. A reviewed subset of settings writes, settings documents, return to measurement, and manual zeros and spans watched at the panel or driven from the host (fuji-calibrate) work on the development analyzer too; auto calibration and auto zero have run only on the simulated analyzer, since the development analyzer has no calibration valves. See Safety. Start with the async quickstart.

The authoritative architectural document is the Design. What the development analyzer actually does on the wire is recorded in Protocol findings.

  • Channel-oriented reads. A full poll of every channel and the analyzer's status takes two Modbus transactions.
  • Validity and provenance on every value. Hold, calibration and error state, and where each gas label came from, travel with each reading.
  • Writes restricted to a frozen envelope of documented user settings and operation commands, behind confirm=True and read-back verification.
  • No hardware needed to develop or test: a simulated analyzer drives the full stack in CI.

Oxygen

Over Modbus, O2 arrives with the display's resolution: 0.01 vol% per step on the development unit. Modbus O2 is not validated for oxygen-consumption calorimetry; see Measurement quality.

Also sold as

ZP-series analyzers are sold under other names too:

Sold as By Status
ZPA NDIR/O2 Multichannel Analyzer (CAI ZPA) California Analytical Instruments (CAI), now part of ENVEA (CAI ENVEA Group) The development analyzer is a CAI-branded ZPA.
IR202 Infrared Gas Analyzer Yokogawa Untested. Its MODBUS manual (IM 11G02Q02-51EN) documents the same register map and the same 38400 8-N-1 link.

fujilib is an independent project. It is not affiliated with or endorsed by Fuji Electric, California Analytical Instruments, ENVEA or Yokogawa; their names and product names belong to their owners.